ProLine PCD
Product Description
The ProLine PCD (Process Challenge Device) contains a paper disc inoculated with a sterilization-resistant spore population. The spore disc is packaged in a glassine envelope and is located in the center of the ProLine housing. The spores on the disc will be deactivated if adequate sterilant penetration occurs.
Model
- PL-3-6-15 (15/box)
Technical Parameters
Organism | Geobacillus stearothermophilus 79531 |
Spore Count | ≥1.0 x 106 |
D-Value | ≥1.5 minutes |
Z-Value | N/A |
Incubation | Time: 7 days Temperature: 55-60˚C Media: SCDB |
Storage | 15-27°C at 30-70%RH. Do not desiccate |
Shelf-Life | 18 months from date of manufacture |
Nominal Dimensions | Disc: 6mm Device length: 130mm Device diameter: 50mm |
1 or equivalent recognized culture collection strain
Indications for Use
ProLine PCD is a Biological Indicator system designed for use in the validation or monitoring of steam sterilization cycles of tubing ranging from 1/8” to 5/8”.
Packaging
Proline PCD comes as a preassembled device in a chipboard box. There are fifteen devices per box.
Labeling
- The plastic device is labeled with lot number and expiration date.
- Each box has a self-adhesive label on the outside to further identify the product.
ProLine PCD
Product Description
The ProLine PCD (Process Challenge Device) contains a paper disc inoculated with a sterilization-resistant spore population. The spore disc is packaged in a glassine envelope and is located in the center of the ProLine housing. The spores on the disc will be deactivated if adequate sterilant penetration occurs.
Model
- PL-3-6-15 (15/box)
Technical Parameters
Organism | Geobacillus stearothermophilus 79531 |
Spore Count | ≥1.0 x 106 |
D-Value | ≥1.5 minutes |
Z-Value | N/A |
Incubation | Time: 7 days Temperature: 55-60˚C Media: SCDB |
Storage | 15-27°C at 30-70%RH. Do not desiccate |
Shelf-Life | 18 months from date of manufacture |
Nominal Dimensions | Disc: 6mm Device length: 130mm Device diameter: 50mm |
1 or equivalent recognized culture collection strain
Indications for Use
ProLine PCD is a Biological Indicator system designed for use in the validation or monitoring of steam sterilization cycles of tubing ranging from 1/8” to 5/8”.
Packaging
Proline PCD comes as a preassembled device in a chipboard box. There are fifteen devices per box.
Labeling
- The plastic device is labeled with lot number and expiration date.
- Each box has a self-adhesive label on the outside to further identify the product.